<5 INSTALLATION AND WIRING> 5-4

IM 11Y01B02-01E-A 4th Edition September 11, 2012-00

6

Factory

1

Ethernet

+ Transmit

2

- Transmit

3

+ Receive

4

- Receive

7

Factory

1-8

Detect

Internal connections only – do not use

8

Field

Analog

#1 output

+

Cal/Val initiate signal loop (SV # 1) to Remote voltage

free contacts/switch. Do not apply external power!

-

Analog

#2 output

+

Cal/Val initiate signal loop (SV # 2) to Remote voltage

free contacts/switch. Do not apply external power!

-

Analog

#3 output

+

Cal/Val initiate signal loop (SV # 2) to Remote voltage

free contacts/switch. Do not apply external power!

-

9

Factory

and/or

Field

1

Gas

Temperature

Compensation

Externally powered 4-20mA gas temperature signals

are wired to 1 (+) and 2(-), 3 not used.

Loop powered temperature transmitter can be con-

nected to 1 (-) and 3 (+24VDC), 2 not used

2
3

4

Gas

Pressure

Compensation

Externally powered 4-20mA gas pressure signals are

wired to 1 (+) and 2(-), 3 not used.

Loop powered pressure transmitter can be connected

to 1 (-) and 3 (+24VDC), 2 not used

5
6

14

Factory

1-4

Mini-Display

Internal connections only – do not use

Figure 9 – Internal Wiring Diagram, including optional cell heating and optional purge unit