<6 BASIC OPERATION> 6-5
IM 11Y01B02-01E-A 4th Edition September 11, 2012-00
6.2 Software Guide
MAIN MENU
Display of Concentration & Transmission
Status Window – notification of initiating, working
properly, warnings or faults
Gas Temperature
Gas Pressure
Selection of Basic or Advanced Menu
Active Alarm Display Button
Analyzer Shut Down Button
Tag number and serial number
Analyzer date and time
After selection of either Basic or Advanced Menu
you will see the Output Selection screen.
This allows control of the analog output while the user
is working in the analyzer software.
•
Block will hold outputs at 3.8mA until
return to Main Screen
•
Track will allow outputs to continue to
report concentration and transmission
until return to Main Screen
•
Hold will hold outputs at their current
value until return to Main Screen
BASIC MENU
Configure – allows setting of Path Length, Gas Tem-
perature, Gas Pressure
View Spectra – user will select display of raw detector
signal or absorption spectra
Data – Alarm History, Calibration History, Record
Result Data
Trends – allows for the displaying of data in a trend
format