Current ramp to laser 

Signal at Detector 

Processed Detector Signal 

 

<2 INTRODUCTION AND GENERAL DESCRIPTION> 2-2

IM 11Y01B02-01E-A 4th Edition September 11, 2012-00

2.1.1 Measurement

• During measurement the laser is held at a fixed temperature.

This is the coarse wavelength adjustment.

• A current ramp is fed to the laser. This is the fine wave

length adjustment.

• The current is ramped to scan across the wavelength

region desired.

• The collimated light passes through the gas to be

measured. The amount of light absorbed by the peak is pro

proportional to the target gas concentration.

• The light is then focused on a detector.

• This signal is used to quantify the light absorbed by the target

gas.

Figure 1- Basic Layout