<8 TROUBLESHOOTING> 8-2

IM 11Y01B02-01E-A 4th Edition September 11, 2012-00

8.1 Common Troubleshooting Steps

For most conditions the troubleshooting steps are common. In general, the most common issues with

the analyzer revolve around ensuring an adequate amount of the laser light is received at the detector.

Under any warning conditions it is recommended that the following steps are carried out prior to

contacting YOKOGAWA.

8.1.1 On Process Gas or Zero Gas or Span Gas

Check Status LEDs.

This will ensure that power is routed properly to the system components.

Status LEDs are listed below:

Blue LED #

(from left to right)

Voltage

Use

D20

+5V

Laser temp control power supply

D26

-15V

Isolated power supply

D27

+15V

Isolated power supply

D24

+8V

DFB laser driver power supply

D23

-8V

DFB laser driver power supply

D22

-12V

DFB laser driver power supply

D16

-12V

Detector board power supply

D19

-12V

VCSEL laser driver power supply

D6

-15V

Analog IO board power supply

D12

-15V

FPGA board power supply

D18

+12V

VCSEL laser driver power supply

D7

+12V

LCD power supply

D8

+24V

Main power supply

D9

+15V

Analog IO board power supply

D10

+5V

SBC power supply

D11

+15V

FPGA board power supply

D13

+5V

FPGA board power supply

D14

+6V

FPGA board power supply

D15

+12V

Detector board power supply

D17

+12V

General 12V power supply