8 troubleshooting> 8-2, 1 common troubleshooting steps, 1 on process gas or zero gas or span gas – Yokogawa Extractive Oxygen Gas Analyzer TDLS220 User Manual
Page 70
<8 TROUBLESHOOTING> 8-2
IM 11Y01B02-01E-A 4th Edition September 11, 2012-00
8.1 Common Troubleshooting Steps
For most conditions the troubleshooting steps are common. In general, the most common issues with
the analyzer revolve around ensuring an adequate amount of the laser light is received at the detector.
Under any warning conditions it is recommended that the following steps are carried out prior to
contacting YOKOGAWA.
8.1.1 On Process Gas or Zero Gas or Span Gas
•
Check Status LEDs.
This will ensure that power is routed properly to the system components.
Status LEDs are listed below:
Blue LED #
(from left to right)
Voltage
Use
D20
+5V
Laser temp control power supply
D26
-15V
Isolated power supply
D27
+15V
Isolated power supply
D24
+8V
DFB laser driver power supply
D23
-8V
DFB laser driver power supply
D22
-12V
DFB laser driver power supply
D16
-12V
Detector board power supply
D19
-12V
VCSEL laser driver power supply
D6
-15V
Analog IO board power supply
D12
-15V
FPGA board power supply
D18
+12V
VCSEL laser driver power supply
D7
+12V
LCD power supply
D8
+24V
Main power supply
D9
+15V
Analog IO board power supply
D10
+5V
SBC power supply
D11
+15V
FPGA board power supply
D13
+5V
FPGA board power supply
D14
+6V
FPGA board power supply
D15
+12V
Detector board power supply
D17
+12V
General 12V power supply