IM 1E8B0-01E
7-13
7. OPERATION VIA BRAIN TERMINAL
7.2.10 Alarm Output (Refer to 6.3.8)
Display
Description
Set "ALARM" in parameter number "B10" to use the P+ and P- terminals for
alarm output only.
Code
Content
PULSE
ALARM
Pulse output
Alarm output
Default
PARAM
B10:PULSE/ALARM
ALARM
B11:4-20 ALM OUT
2.4mA OR LESS
B12:POWER FREQ
50.03Hz
DATA
ESC
DIAG
DATA
ESC
DIAG
* The "N" item can be opend by entering "55" in parameter number "L02".
7.2.11 Data Setting Enable / Inhibit (Refer to 6.3.9)
• This function makes it possible to inhibit all data changes (except L1). However,
automatic zero adjustment can be made if the automatic zero-adjustment function is
enabled (set in C1). Also, totalization presetting can be done if the totalization
presetting function is enabled (set in E4).
Display
Description
Selecting "INHIBIT" in parameter number "L01" disables data setting.
Code
Content
INHIBIT
ENABLE
Data setting inhibit.
Data setting enabled.
Default
PARAM
L01:TUNING
INHIBIT
L02:KEY
00
L60:SELF CHECK
GOOD
DATA
ESC
DIAG
DATA
ESC
DIAG
7.2.12 Procedure of Selecting Special Application Items (Refer to
6.3.10)
Display
Description
It is possible to open up to item N when "55" is entered in parameter number
"L02."
Code
Content
00
55
Accessible up to L parameters
Accessible up to N parameters
Default
PARAM
L01:TUNING
INHIBIT
L02:KEY
55
L60:SELF CHECK
GOOD
DATA
ESC
DIAG
DATA
ESC
DIAG