4.INSTALLATION
I
M 1E8B0-01E
4-14
4.4 Wiring
Confirm that all connections are correct before applying power to the
instrument. Improper wiring may damage the flowmeter.
The external signal wirings are connected into the terminal inside the converter.
Please connect to each terminal (Please refer to Figure 4.4.1) by taking off a cover
backside the converter.
Power supply
Protective grounding
4 to 20 mA DC output
Pulse or alarm
Terminal Symbols
Description
N–
L+
I+
I–
P+
P–
Protective grounding
Current Output 4 to 20mA DC
Power Supply
Pulse or alarm output
Figure 4.4.1 Terminal
Please give attention to avoid the cable is bended excessively.
4.4.1 Protective Grounding
Please be sure to connect protective grounding of ADMAG CA with cable of
2mm
2
or larger cross section in order to avoid the electrical shock to the
operators and maintenance engineers and prevent the influence of external
noise. And further connect the grounding wire to the mark (100
Ω
or
less).
CAUTION
CAUTION