beautypg.com

Yokogawa ADMAG CA User Manual

Page 36

background image

IM 1E8B0-01E

5-7

5. BASIC OPERATING PROCEDURES

(4) Display of 5-digit Data

• The data display area has four digits and

can therefore not display span and other 5-
digit data parameters.
To display such data, the data is automati-
cally shifted (scrolled) to the left one digit
at a time .
When a 5-digit parameter is selected, the
delimiter " : " starts to flash.

• When the flashing segment is aligned with

the last digit and the data includes a space,
the data is displayed as shown “4” on the
right.

(5) Display of 6-digit Data

• The data display area has four digits and can therefore not display span and other 6-

digit data parameters. (E05 TL SET VALUE is the only 6-digit parameter.)
Like 5-digit data, 6-digit data is displayed by automatically shifting (scrolling) it to
the left one digit at a time.

(6) Display and Setting Coded Data

• Example of change : Changing from +120 to -120

Item

Display

Content

SET SHIFT INC

SET SHIFT INC

SET SHIFT INC

The display on right indicates "+120".

Cause the "+" sign to flash.

Press the INC key to change the "+" sign to a "-" sign.

Press the SET key twic.

1. Selecting
Coded Data

2. Coded Data
Flashes

3. Change of
Coded Data

1

1 second

1 second

1 second

1 second

1 second

1 second

(Space)

(Space)

3 seconds

Display example : Display of span 123.45

2

3

4

5

6

7