IM 1E8B0-01E

13-2

(4) Maintenance and Repair

The instrument modification or parts replacement by other than authorized
representative of Yokogawa Electric Corporation is prohibited and will void
the approval of Factory Mutual Research Corporation.

HAZARDOUS LOCATIONS

Magnetic Flow Converter

Conduit

Sealing Fitting

18"(457mm) Max.

(Note) In the case of non incendive for Division 2 (Zone 2), sealing fittings are not

required and installation shall be in accordance with the National Electric
Code, ANSI/NFPA-70.

13.2 CSA (Only for sizes 15 to 100 mm(0.5 to 4 in))

(1) Technical Data

• Applicable Standard:

C22.2 No 0, C22.2 No 0.4, C22.2 No 0.5, C22.2 No 25, C22.2 No 30,
C22.2 No 94, C22.2 No 142, C22.2 No 157, C22.2 No 213, C22.2 No 1010.1

• Certificate: 1500865
• Explosion proof for Class I, Groups B, C and D; Class III, Groups E, F and G; Class III;

Temp. Code:

T6

T5

T4

Process Temp.:

70

85

120˚C

Ambient Temp.: –20

°

C to +50

°

C (–4 to 122

°

F)

Maximum power supply voltage: 250 Vac/ 130 Vdc
Enclosure: Type 4X

• Non incendive for Class I, Div.2 Groups A, B, C and D; Class II, Div.2, Groups E, F

and G; Class III:

Temp.Code: T4A
Process Temp.: 105˚C
Ambient Temp.: –20

°

C to +50

°

C (–4 to 122

°

F)

Maximum power supply voltage: 250 Vac/ 130 Vdc
Enclosure: Type 4X

(2) Installation

All wiring shall comply with Canadian Electrical Code Part I and Local
Electrical Codes.
In hazardous location, wiring shall be in conduit as shown in the figure.

CAUTION :

SEAL ALL CONDUITS WITHN 50cm OF THE ENCLOSURE’

UN SCELLEMENT DOIT ÉTRE INSTALLÉ À MOINS DE

50cm DU BOÎTIER.