8 reversing flow direction (refer to 6.3.6), 9 limiting current output (refer to 6.3.7), Reversing flow direction (refer to 6.3.6) -12 – Yokogawa ADMAG CA User Manual
Page 62: Limiting current output (refer to 6.3.7) -12
IM 1E8B0-01E
7-12
7. OPERATION VIA BRAIN TERMINAL
7.2.8
Reversing Flow Direction (Refer to 6.3.6)
Display
Description
The flow direction can be set in parameter number “B14”.
Code
Description
FORWARD
REVERSE
Forward direction
Reverse direction to flow arrow
PARAM
B13:VELOCITY CHK
5 m/s
B14:FLOW DIR
REVERSE
B60:SELF CHECK
GOOD
DATA
ESC
DIAG
7.2.9
Limiting Current Output (Refer to 6.3.7)
(1) 4 to 20 mA Low Cut Output (Current Output Near by 0% Range)
Display
Description
PARAM
G01:4-20 LOW CUT
10 %
G02:4-20 LOW LMT
-20 %
G03:4-20 H LMT
110 %
DATA
ESC
DIAG
Setting range : 0 to 10%
Default: 0%
0.5% 0.5%
4mA
5.6mA(10%)
Output
Set value
Hysteresis: 1% fixed
(2) 4 to 20 mA Low Limit
Display
Description
Setting range: -20 to 100%
Default: -20%
4mA
5.6mA(10%)
Output
DATA
ESC
DIAG
PARAM
G01:4-20 LOW CUT
0 %
G02:4-20 LOW LMT
10 %
G03:4-20 H LMT
110 %
(3) 4 to 20 mA High Limit
Display
Description
Setting range: 0 to 110%
Default: 110%
18.4mA(90%)
4mA
Output
PARAM
G01:4-20 LOW CUT
0 %
G02:4-20 LOW LMT
-20 %
G03:4-20 H LMT
90 %
DATA
ESC
DIAG