IM 1E8B0-01E

7-8

7. OPERATION VIA BRAIN TERMINAL

7.2.2

Measuring Mode Setting

Measuring Modes

The measuring mode is Standard or Noise Reduction mode. Standard mode is set before
shipment. If the output fluctuation is large when a fluid actually flows, the fluctuation is
reduced by increasing the damping time constant.

• If the output fluctuation is still large when the damping time constant is increased, use

the Noise Reduction mode.

• Use the standard mode for sticky fluids.

After the measuring mode is changed from Standard mode to Noise Reduction mode or
from Noise Reduction mode to Standard mode, be sure to adjust the zero point. Perform
zero adjustment with the metering tube filled with the fluid and with the fluid static.

• Standard mode (STD) must be set in case of size 150, 200 mm (6, 8 in).

It cannot be measureed in Noise Reduction mode (FNR).

Display

Description

PARAM
B13:VELOCITY CHK

1.00 m/s

B14:FLOW DIR

FORWARD

B15:MEAS MODE

STD

DATA

ESC

DIAG

Select the measuring mode in parameter number “B15”.

Code

Measuring Mode

STD

FNR

Standard mode
Noise Reduction mode

IMPORTANT