REPLACE BATTERY

Battery voltage is less than absolute minimum. Shutdown
imminent.

Replace with charged battery.

REPLACE CARD

Write errors during manual or semi-automated modes.

May have configuration card installed or write
protection on.

REPORT FULL

Summary report memory full.

Erase summary report.

REPORT HALTED

Summary report stops printing unexpectedly.

Turn unit off and then back on again. Print
Summary again. If fault persists, replace system
board.

SELECT 30J FOR TEST

Attempt to run a self test at an energy other than 30J.

SELECT DEFIB MODE

Analyze button pressed in pace or monitor mode.

SELECT LEADS

12 Lead monitor determined the user selected MFE.

SELECT LIMB LEADS

Paddles or augmented ECG leads selected when
continuous analysis active or started.

Select limb leads I, II, III or MFE

SELECT PADS

Lead I, II, or III selected when analyze pressed.

SET CLOCK

Real time clock failure: invalid date or time.

Set date and time and/or replace system board.

SET PACE mA

Multiple copy errors are the product of intended software
or memory errors. If error reoccurs other than on entering
pace the first time or after more than 10 minutes in other
mode, the unit could be broken.

Set pace current. If broken, replace system board.

SET PACE RATE

Multiple copy errors are the product of intended software
or memory errors. Multiple copies of pace rate don't
match. If error persists, unit could be broken

Set pace rate. If broken, replace system board.

SHOCK ADVISED

Advisory message when analysis finds a shockable
rhythm.

SpO

2

AMBIENT LIGHT

Ambient light is too bright.

Shield sensor from ambient light. Replace Sp0

2

sensor. Replace Sp0

2

module