ZOLL E Series Monitor Defibrillator Rev D User Manual
Page 64
E Series Service Manual
2-12
9650-1450-01 Rev. D
REPLACE BATTERY
Battery voltage is less than absolute minimum. Shutdown
imminent.
Replace with charged battery.
REPLACE CARD
Write errors during manual or semi-automated modes.
May have configuration card installed or write
protection on.
REPORT FULL
Summary report memory full.
Erase summary report.
REPORT HALTED
Summary report stops printing unexpectedly.
Turn unit off and then back on again. Print
Summary again. If fault persists, replace system
board.
SELECT 30J FOR TEST
Attempt to run a self test at an energy other than 30J.
SELECT DEFIB MODE
Analyze button pressed in pace or monitor mode.
SELECT LEADS
12 Lead monitor determined the user selected MFE.
SELECT LIMB LEADS
Paddles or augmented ECG leads selected when
continuous analysis active or started.
Select limb leads I, II, III or MFE
SELECT PADS
Lead I, II, or III selected when analyze pressed.
SET CLOCK
Real time clock failure: invalid date or time.
Set date and time and/or replace system board.
SET PACE mA
Multiple copy errors are the product of intended software
or memory errors. If error reoccurs other than on entering
pace the first time or after more than 10 minutes in other
mode, the unit could be broken.
Set pace current. If broken, replace system board.
SET PACE RATE
Multiple copy errors are the product of intended software
or memory errors. Multiple copies of pace rate don't
match. If error persists, unit could be broken
Set pace rate. If broken, replace system board.
SHOCK ADVISED
Advisory message when analysis finds a shockable
rhythm.
SpO
2
AMBIENT LIGHT
Ambient light is too bright.
Shield sensor from ambient light. Replace Sp0
2
sensor. Replace Sp0
2
module
Error Message
Explanation
User
Advisory
Technical Action