Defibrillator – ZOLL E Series Monitor Defibrillator Rev D User Manual
Page 243
Troubleshooting Guidelines
9650-1210-01 Rev. R
13-5
Defibrillator
Symptom
Recommended Action
Excessive artifact when using paddles as ECG source.
•
Ensure PADDLES is selected.
•
Firmly press paddles against patient skin.
•
Use gel on paddles.
•
Clean paddle surface.
•
Check and clean between adult and pediatric shoe.
•
Check cable for damage.
•
Use MFE electrodes for ECG analysis.
Defibrillator will not charge (energy level does not
increment on display).
•
Check SHOCK button(s) on paddles or front panel are
not stuck on.
•
Replace battery with a fully charged battery pack.
Charge time to 200 J exceeds 10 seconds.
•
Typical in a low battery condition (up to 20 seconds)
•
Change battery pack.
•
Plug device into AC power.
•
Install fully charged battery pack.
Energy does not discharge when you press SHOCK
button(s).
•
60 seconds have elapsed in Manual mode since initial
charge ready. Energy was internally discharged.
•
15 seconds have elapsed in automatic mode since
initial charge ready. Energy was internally discharged.
•
Device is in SYNC mode and no QRS complex is
detected.
•
Energy internally discharged because energy selection
was changed during charge or after the device was
ready.
•
Unit not completely charged when you pressed SHOCK
button(s). Wait for DEFIB XXXJ READY message and
ready tone.
•
Press and hold SHOCK button(s) until energy is
delivered to the patient.
•
Device is connected to the AutoPulse Plus and no
relaxation period of the compression cycle has been
detected yet.
Unit is unable to SHOCK when in SYNC mode.
•
Ensure SYNC XXXJ SEL message is displayed on
monitor.
•
Check for SYNC marker above R wave. If not present,
change ECG size, lead selection, or electrode
placement.
•
Press and hold SHOCK button(s) until energy is
delivered to the patient.
•
Alter ECG electrode placement.
•
Make sure ECG signals are displayed.