E Series Service Manual

9650-1450-01 Rev. D

5-7

6. Flip front panel over. Remove main selector knob by gently

lifting. (Note: Older version may be glued and need to be pried
with cutter.)

7. Remove 1/2” nut from the main selector knob, then remove

the anti-rotation washers.

8. (Unit with pace option) Remove the pacer controls knobs, and

then remove 1/2” nut and washer.

9. Flip front panel over. Disconnect front membrane cable and

data entry switch from control board.