ZOLL E Series Monitor Defibrillator Rev D User Manual
Page 63
![background image](/manuals/355809/63/background.png)
E Series Service Manual
9650-1450-01 Rev. D
2-11
PADDLE FAULT
Cannot detect type of accessory attached to the universal
cable.
Replace paddles, internal paddles, system board,
high voltage module and/or universal cable.
PERFORM CPR
Advisory message in AED auto defib mode.
POOR LEAD CONTACT
One or more ECG leads are poorly connected or not
connected to patient. (User configurable.)
Check electrode attachment to patient, cable
connector to electrode, cable to unit connector.
Broken unit.
POOR PAD CONTACT
Electrode impedance exceeds threshold.
Ensure pads are coupled to patient. Check /
replace pads or universal cable.
Check impedance circuit calibration.
Replace system board.
PRESS ANALYZE
Alternate message for check patient prompt.
PRESS CHARGE
Advisory message in conjunction with shock advised.
PRESS SHOCK
Prompt issued in AED auto defib mode when defib is
charged (ready).
RECORDER FAULT 142
Strip chart system error.
Check paper tray and paper path. Replace the
print head, system interconnect board and or the
system board.
RECORDER FAULT 143
Strip chart failed power-up echo test. Communications
error.
Check paper tray and paper path. Replace the
system interconnect board and/or the system
board. Turn unit off and back on again.
RECORDER FAULT 147
Strip chart print head over temperature.
Check paper tray and paper path. Replace the
print head, system interconnect board and/or the
system board.
RELEASE BUTTONS
Simultaneous external paddle button presses detected
before unit reached full defib charge (ready state).
Release buttons.
RELEASE SHOCK
Discharge switch(es) closed when pressing charge button.
Discharge button pressed before defib reached ready
state.
Release shock button.
Check paddles.
Replace controls board.
Error Message
Explanation
User
Advisory
Technical Action