E Series Operator’s Guide
12-4
9650-1210-01 Rev. R
•
DO exchange your battery packs regularly.
Battery packs should be exchanged once per shift or once per day depending on their use.
•
DO carry a fully charged spare battery pack at all times.
DON’T remove a partially charged battery pack from the battery charger.
If a partially charged battery pack is used, a full charge is recommended before its next use. Repeated use after
partial charging may quickly diminish the battery’s capacity, thereby shortening its life.
DON’T store battery packs in a fully discharged state.
Battery pack capacity diminishes if left in a fully discharged state for extended periods.
DON’T assume that a shift check of the monitor/defibrillator verifies adequate battery pack run time.
Your Monitor/Defibrillator should be tested daily to verify the readiness of the device. This test, however, does not
verify adequate charge state or capacity of the battery pack and may leave the monitor/defibrillator with inadequate
run time.
If the device shows a LOW BATTERY warning during testing, replace the depleted battery pack with a fully charged
one, and recharge the depleted battery pack.
DON’T charge battery packs at temperature extremes.
ZOLL recommends charging battery packs at or near normal room temperature (15°C to 35°C or 59°F to 95°F).