9650-1210-01 Rev. R
A-5
Display
Recorder
PCMCIA Card
Battery Packs
Screen Type
High resolution, color liquid crystal display (LCD)
Screen Size
5.63 inches (14.3 cm) diagonally
Sweep Speed
25 mm/s
Viewing Time
4 seconds
Messages
ERASING REPORT, INSERT CARD, CARD FULL, REPLACE BATTERY, LOW BATTERY,
PERFORM CPR, ECG TOO LARGE, NOISY ECG, RETRY ANALYSIS, CHECK PATIENT,
ANALYSIS HALTED, PRESS ANALYZE, NO SHOCK ADV., CHECK PULSE,
SHOCK ADVISED, PRESS CHARGE, SELECT PADS, SELECT ECG LEADS,
SELECT DEFIB MODE, VF ALARMS OFF, DISABLE SYNC, ANALYSIS RESTARTED,
CHECK PADS, POOR PAD CONTACT, DEFIB PAD SHORT, PADDLE FAULT,
ECG LEAD OFF, USE PADDLE DISCHG, OPEN AIR DISCHARGE, CANNOT CHARGE,
RELEASE SHOCK, PRESS SHOCK, TEST OK, TEST FAILED, PACER DISABLED,
DEFIB DISABLED, SET PACE MA, SET PACE RATE, CHECK RECORDER,
ANALYZING ECG, FULLY RELEASE.
Paper
80 mm thermal (grid width)
90 mm (paper width)
Speed
25 mm/s or 50 mm/s (apparent), configurable for 12 lead output
Delay
6 seconds
Annotations
Time, date, defib energy, heat rate, pacer output (Pacer version only), QRS sync marker,
ECG SIZE, lead, alarm, DEFIB TEST OK/FAIL, ANALYZE ECG, PADS OFF, ANALYSIS
HALTED, NOISY ECG, SHOCK ADVISED, NO SHOCK ADVISED, ECG TOO LARGE, and
diagnostic bandwidth.
Printing Method
High resolution, thermal array print head
Print out Modes
Manual or automatic; user-configurable
On/Off Control
Front panel and paddle
Automatic Function
15 second recording initiated by alarm activation or defibrillator discharge
Capacity
Standard Series II Flash card - 8, 16, or 32 Mb
Audio Recording
Digital compressed audio data
Type
Rechargeable, sealed lead acid
Lithium ion
Weight
1 kg (2.2 lbs)
0.68 kg (1.5 lbs)
Voltage
2 Volts per cell; 5 cells wired in series
4.2 Volts per cell; 3 cells wired in series
Recharge Time
(to 90% capacity)
4 hours or less with integral charger
7 hours or less with integral charger