ZOLL E Series Monitor Defibrillator Rev D User Manual
Page 61
E Series Service Manual
9650-1450-01 Rev. D
2-9
ECG LEAD OFF
One or more ECG leads are not properly connected when
leads are selected as input.
Check cable and patient connection. Change
electrodes. Prepare patient’s skin.
ECG TOO LARGE
ECG signal too large for accurate shockable rhythm
analysis.
Reduce ECG size.
ECG V LEAD OFF
V LEAD cable is not properly attached.
Attach V LEAD cable.
ECG V1 LEAD OFF
Chest lead V1 is not properly attached to patient.
Reattach V lead. Check cable.
ECG V2 LEAD OFF
Chest lead V2 is not properly attached to patient.
Reattach V lead.
Check cable.
ECG V3 LEAD OFF
Chest lead V3 is not properly attached to patient.
Reattach V lead.
Check cable.
ECG V4 LEAD OFF
Chest lead V4 is not properly attached to patient.
Reattach V lead.
Check cable.
ECG V5 LEAD OFF
Chest lead V5 is not properly attached to patient.
Reattach V lead.
Check cable.
ECG V6 LEAD OFF
Chest lead V6 is not properly attached to patient.
Reattach V lead. Check cable.
ENTER ACCESS CODE
Manual mode access code needed.
Enter access code to enter manual mode with
AED.
ERASING REPORT
Summary report being erased.
FAX BUSY
FAX line at receiving FAX is busy.
FAX DIALING
Preparation for sending fax.
FAX DONE
Transmission complete.
FAX PREPARING
Preparing fax for transmission.
FAX SENDING
Transmitting fax.
Error Message
Explanation
User
Advisory
Technical Action