Sync 200j sel, Caution, Warning – ZOLL E Series Monitor Defibrillator Rev D User Manual
Page 202
E Series Operator’s Guide
6-2
9650-1210-01 Rev. R
Turn Selector Switch to MONITOR
Press the LEAD button to select the desired ECG lead.
The lead you select is then displayed at the top of the
screen.
1
Select DEFIB
Turn the selector switch to DEFIB, then select the
desired energy level using the up/down arrows on the
front panel or sternum paddle (if being used).
Press SYNC softkey
The selected energy level is displayed on the monitor.
A SYNC marker (
) appears on the monitor above each
detected R-wave to indicate where discharge will occur.
Verify that markers are clearly visible on the monitor and
their location is appropriate and consistent from beat to
beat. If necessary, use the LEAD button and SIZE button
to establish settings that yield the best display.
•
A SYNC XXXJ SEL. message appears on the
display. If the DEFIB XXXJ SEL. message appears,
press the SYNC softkey. Two quick beeps sound
when the SYNC softkey is pushed.
Unless otherwise configured, Sync mode is turned off
automatically after each shock or if the selector switch
has been moved to PACER or OFF.
Press the SYNC softkey again to reactivate SYNC
mode. Changing the selected energy does not turn off
Sync mode.
You can configure the unit to stay in Sync mode after
defibrillation if desired.
2
Charge Defibrillator
Press the CHARGE button on the front panel or on the
apex paddle handle (if paddles are used).
To abort charging and increase or decrease the selected
energy after you press the CHARGE button, use the
ENERGY SELECT buttons on either the sternum paddle
(if paddles are used) or defibrillator front panel. Press the
CHARGE button again to charge the unit.
After charging to the selected energy, the SHOCK button
on the front panel lights. If paddles are used, the charge
indicator on the apex paddle illuminates. A distinctive
audible tone sounds and the energy ready SYNC XXXJ
READY message is displayed.
The defibrillator is now ready.
3
Deliver SHOCK
Verify that the ECG waveform is stable and that a marker
appears only with each R-wave.
Press and hold the illuminated, front panel SHOCK
button or, if paddles are used, simultaneously press and
hold both SHOCK buttons (one on each paddle) until
energy is delivered to the patient. The defibrillator
discharges with the next detected R-wave.
0!#%2
/&&
-/.)4/2
$%&)"
3(/#+
!.!,9:%
#(!2'%
%.%2'9
3%,%#4
0!#%2
/&&
-/.)4/2
$%&)"
and
SYNC 200J SEL.
ID#
Param
00:01
ECG
Wave 2
Alar ms
Sync
On/Off
CAUTION
Changing the selected energy while the unit is
charging or charged will cause the defibrillator to
disarm itself. Press the CHARGE button again to
charge the unit.
WARNING
•
Warn all persons in attendance of the patient to
STAND CLEAR prior to defibrillator discharge.
•
Verify that no one is in contact with the patient,
monitoring cable or leads, bed rails, or any other
potential current pathways.
3(/#+
!.!,9:%
#(!2'%
%.%2'9
3%,%#4
or