ZOLL E Series Monitor Defibrillator Rev D User Manual
Page 179
Operating Controls and Indicators
9650-1210-01 Rev. R
2-7
Heart Rate Alarm Activated Format
The summary report function records 6 seconds of pre-alarm patient ECG. Also recorded are the ECG lead, ECG size,
patient's heart rate, actual event time, and date. The date/time printed on the top of the strip corresponds to the ECG
data occurring 6 seconds before the event. If the pacer is on during this event, the pacing rate and pacing current are
also recorded.
VF Alarm Activated Format (Refer to Section 6)
The summary report function records 18 seconds of patient ECG data associated with each VF alarm. Also recorded
are the shock count, ECG lead, ECG size, actual event time, patient’s heart rate, and noise events. The date/time
printed on the top of the strip corresponds to the ECG data occurring 6 seconds before the event.
Recorder On Format
The summary report function records 6 seconds of patient ECG prior to turning on the recorder. Also recorded are the
ECG lead, ECG size, patient's heart rate, actual event time, and date. The date/time printed on the top of the strip
corresponds to the ECG data occurring 6 seconds before the event. If the pacer is on during this event, the pacing rate
Actual
event
date/time
Actual
event
date/time