beautypg.com

Required equipment, Safety precautions, Required equipment -2 – ZOLL E Series Monitor Defibrillator Rev D User Manual

Page 112: Safety precautions -2

background image

E Series Service Manual

5-2

9650-1450-01 Rev. D

Required Equipment

#1 Philips head screwdriver

#2 Philips head screwdriver

#24 Hex head screwdriver

Wooden stick

Needlenose pliers

Wire Cutters

Exactor Knife

1/2” Nut Driver

Safety Precautions

Warning!

SHOCK HAZARD!

Caution

TAKE THE NECESSARY PRECAUTIONS TO GUARD AGAINST SHOCK OR INJURY BEFORE YOU CONDUCT
DEFIBRILLATOR TESTS OR REPAIRS.

Only properly trained technicians should service the unit.

The unit can contain deadly voltages even if the unit is turned off.

Make sure to discharge the unit before working with it.

Make sure you take the necessary precautions when working with static sensitive units. For example, you must wear a conductive wrist strap (which
touches your skin) connected to a grounding mat and to the earth ground. You must remove the wrist strap when you discharge high voltage or when
you are working on energized equipment.

Wear gloves to prevent skin oils from affecting the equipment.