Maintenance tests, Overview, Chapter 1 – ZOLL E Series Monitor Defibrillator Rev D User Manual
Page 11: Overview -1, Chapter 1 maintenance tests
E Series Service Manual
9650-1450-01 Rev. D
1-1
Chapter 1
Maintenance Tests
Overview
The E Series has two checkout procedures: the operator’s shift checklist and the extensive six-month maintenance tests checkout procedures.
Because the E Series units must be maintained ready for immediate use, it is important for users to conduct the Operator’s Shift Checklist procedure at the
beginning of every shift. This procedure can be completed in a few minutes and requires no additional test equipment. (See the ZOLL E Series Operator’s
Guide for the Operator’s Shift Checklist.)
A qualified biomedical technician must perform a more thorough maintenance test checkout every six months to ensure that the functions of the E Series
unit work properly. This chapter describes the step by step procedures for performing the six month maintenance test checkout. Use the checklist at the
back of this document (ZOLL E Series Maintenance Tests Checklist) to record your results of the maintenance tests.
This chapter describes the following E Series maintenance tests:
1.
Physical Inspection of the Unit
2.
Front Panel Button
3.
3, 5, and 12 Leads
4.
Power Supply
5.
Leakage Current
6.
Paddles
7.
Heart Rate Display
8.
Calibrating Pulses on Strip Chart
9.
Notch Filter
10.
Heart Rate Alarm
11.
Defibrillator Self Test
12.
Synchronized Cardioversion
13.
Shock