GPS Synchronization

1. Press and hold the left-most softkey on the unit while

turning the selector switch to MONITOR (ON for AED
units). When the System Utilities screen appears on
the monitor, release the softkey.

2. Press the MORE softkey and then the Clock Sync

softkey to display the set time option menu.

3. Press the GPS Time Set softkey to display the GPS

Clock Synchronization screen.

When the GPS time has been acquired, the display
shows the acquired date, time, day and the number of
seconds required to sync the clock. Latitude and
longitude may also be displayed if a complete satellite fix
was obtained. If a partial satellite fix was used to obtain
the clock information, the latitude and longitude will not
be displayed. Press the Return softkey to store the
acquired time and return to the Main System Utilities
screen.

Pressing the Abort softkey cancels the GPS
synchronization; the message GPS ABANDONED is

displayed. Press the Return softkey to return to the Main
System Utilities screen.

Note: GPS synchronization works best when the unit

has a direct line of sight to GPS satellites. Placing
the unit outdoors or adjacent to an unblocked
window when performing GPS synchronization is
advised.

BUSY

The unit detected a busy signal from
the selected NIST location. Retry.

NO ANSWER

The unit received no answer from the
selected NIST location. Retry or
select another NIST location.

NO CARRIER

The unit determined that the line is
broken. Correct the line break and
retry.

HANG UP

The unit received a hang up indication
from the selected NIST location.
Retry.

MODEM DIAL
ERROR

The modem could not dial the phone
number. Ensure that the modem card
is properly connected. Ensure that the
user-selected dial prefix is correct.
(Refer to E Series Configuration
Guide
for information). Retry.

NIST DATA
ERROR

The unit detected an error in the data
from NIST. Retry.

Error Message

Description/Corrective Action

Clock Synchronization

GPS Acquire xxx

Abort

Return