Defib 200j sel, Analyzing ecg, No shock adv – ZOLL E Series Monitor Defibrillator Rev D User Manual
Page 190
E Series Operator’s Guide
4-2
9650-1210-01
Rev.
R
1
Select DEFIB
The unit displays the DEFIB 120J SEL message on the
monitor until you press the ANALYZE button.
Energy Select
Shock 1, Shock 2, and Shock 3 are set to the default
adult settings of 120 J, 150 J, and 200 J respectively.
If medical protocols allow, you may select a different
energy level using the ENERGY SELECT buttons. The
new energy setting is displayed on the monitor.
Automatic energy escalation is disabled if you manually
change the energy level outside the pre-programmed
Shock 1, Shock 2, Shock 3 sequence and deliver a
shock. See the “Energy Level” sections (for Shock 1,
Shock 2, and Shock 3) of the E Series Configuration
Guide for more details.
2
Press ANALYZE Button
Press the ANALYZE button to begin analysis of the
patient’s ECG rhythm and to detect the presence of any
shockable rhythms.
An ANALYZING ECG message is displayed for
approximately 9 to 12 seconds while the patient’s ECG is
analyzed.
Once the analysis is complete, the unit indicates whether
or not a shock is advised.
When a non-shockable rhythm is detected, the message
NO SHOCK ADV. is displayed.
In this case, you should follow local protocols to continue
CPR or other cardiopulmonary life support and re-
analyze the ECG at appropriate intervals.
When a shockable rhythm is detected (ventricular
fibrillation or tachycardia with heart rate > 150):
•
Units with the automatic charge option enabled
automatically charge the defibrillator to the
preconfigured or user selected energy setting.
•
Units with the automatic charge option disabled
alternately display the messages SHOCK ADVISED
and PRESS CHARGE. In this case, charge the
defibrillator by pressing the CHARGE button.
WARNING
•
A patient must be motionless during ECG analysis.
Do not touch the patient during analysis. Cease all
movement via stretcher or vehicle before analyzing
the ECG.
•
When using the AutoPulse Plus, stop
compressions prior to performing ECG analysis.
Compressions may be resumed following the
analysis.
0!#%2
/&&
-/.)4/2
$%&)"
DEFIB 200J SEL.
ID#
Param
00:01
ECG
Wave 2
Alar ms
Sync
On/Off
ANALYZING ECG
ID#
Param
00:01
ECG
Wave 2
Alar ms
Sync
On/Off
DEFIB 200J SEL.
ID#
Param
00:01
ECG
Wave 2
Alar ms
Sync
On/Off
NO SHOCK ADV.