Pacer (Pacer version only)
Symptom
Recommended Action
Unit displays the CHECK PADS prompt.
•
Ensure MFE Pads are connected to Multi-Function
cable.
•
Ensure electrode gel is not dry. Replace MFE Pads if
necessary.
•
Ensure good electrode-to-patient contact.
•
Check integrity of Multi-Function cable by plugging into
test connector. CHECK PADS prompt should disappear.
No pacing marker (
) present on the ECG trace.
•
Ensure unit is in PACER mode.
•
Ensure PACER RATE (ppm) is set greater than patient
heart rate.
No ventricular capture beat after pacing marker is seen
on ECG display.
•
Check patient’s pulse.
•
Increase output current (mA).
•
Ensure MFE Pads are making good contact with the
patient, or review MFE pad placement.
•
Select different ECG lead configuration.
Patient receives intermittent pacing stimulus during
standby pacing.
•
Ensure proper ECG electrode connection and
placement. If ECG lead wire comes off, pacer
automatically paces asynchronously.
•
Check ECG cable for damage.
•
Patient R-to-R interval varying. Pace rate close to
patient rate. Verify rate is set appropriately.
Heart rate display reads 0 with proper pacing capture
displayed on ECG trace.
•
Check patient’s pulse.
•
Select different ECG Lead configuration.
Bedside/Central Station/Telemetry ECG display
becomes erratic when pacing.
•
Patient monitor ECG inputs overloaded by pacer
signals. ECG can only be monitored by the device while
pacing.