Pacer (Pacer version only)

Symptom

Recommended Action

Unit displays the CHECK PADS prompt.

Ensure MFE Pads are connected to Multi-Function
cable.

Ensure electrode gel is not dry. Replace MFE Pads if
necessary.

Ensure good electrode-to-patient contact.

Check integrity of Multi-Function cable by plugging into
test connector. CHECK PADS prompt should disappear.

No pacing marker (

) present on the ECG trace.

Ensure unit is in PACER mode.

Ensure PACER RATE (ppm) is set greater than patient
heart rate.

No ventricular capture beat after pacing marker is seen
on ECG display.

Check patient’s pulse.

Increase output current (mA).

Ensure MFE Pads are making good contact with the
patient, or review MFE pad placement.

Select different ECG lead configuration.

Patient receives intermittent pacing stimulus during
standby pacing.

Ensure proper ECG electrode connection and
placement. If ECG lead wire comes off, pacer
automatically paces asynchronously.

Check ECG cable for damage.

Patient R-to-R interval varying. Pace rate close to
patient rate. Verify rate is set appropriately.

Heart rate display reads 0 with proper pacing capture
displayed on ECG trace.

Check patient’s pulse.

Select different ECG Lead configuration.

Bedside/Central Station/Telemetry ECG display
becomes erratic when pacing.

Patient monitor ECG inputs overloaded by pacer
signals. ECG can only be monitored by the device while
pacing.