E Series Service Manual

2-6

9650-1450-01 Rev. D

CABLE FAULT

(Auto defib mode only.)

Incorrect A/D reading for paddle ID (similar to PADDLE
FAULT).

Replace paddle set, universal cable and/or
system board.

CARD FULL

Memory Card Full.

CHECK CO

2

SENSOR

EtCO

2

Sensor is unplugged or defective.

Check that sensor cable is plugged in and seated
properly. Check that sensor is not exposed to
excessive heat. If problem persists, replace the
sensor.

CHECK CO

2

ADAPTER

Airway adapter is removed, occluded or adapter zeroing
needs to be performed or was performed incorrectly.

Replace/Clean airway adapter. Zeroing performed
automatically.

CHECK PADS

Message displayed in conjunction with either POOR PAD
CONTACT or DEFIB PAD SHORT.

Ensure pads are coupled to patient. Check /
replace pads and universal cable.

Replace system board.

CHECK PATIENT

Background ECG analysis detects shockable rhythm.

CHECK PULSE

Alternate message for NO SHOCK ADVISED
message.Message also shown after delivering third shock
when auto analyze 3 times option is enabled.

CHECK RECORDER

Produced when paper tray is empty, paper jams or
recorder door is opened.

Replace paper sensor board, system interconnect
board, and/or system board.

CHECK SPO

2

SENSOR

Reposition SpO

2

sensor on patient.

CLOCK FAULT 11

Real time clock oscillator failure.

Replace system board.

CLOCK FAULT 12

Real time clock back-up power supply failure. Found
oscillator stopped at power-up, but oscillator now running
when the system is running. (Oscillator only runs when
main power is applied).

Replace system board.

CLOCK FAULT 13

One of the set time units (seconds, minutes, year, etc.) is
out of range.

Replace system board.

Error Message

Explanation

User
Advisory

Technical Action