beautypg.com

Pacemaker (pacer version) defibrillator – ZOLL E Series Monitor Defibrillator Rev D User Manual

Page 248

background image

E Series Operator’s Guide

A-2

9650-1210-01 Rev. R

Pacemaker (Pacer version)

Defibrillator

Type

VVI demand; asynchronous (fixed rate) when used without ECG leads or in ASYNC pacing
mode

Pulse Type

Rectilinear, constant current

Pulse Shape and
Duration

Rectilinear, 40 milliseconds ±2 milliseconds

Pulse Amplitude

Variable 0 mA to 140 mA ±5% or 5 mA, whichever is greater
Digitally displayed on the monitor (increments or decrements by a value of 2 mA)

Pacing Rate

Variable from 30 ppm to 180 ppm ±1.5% (increments or decrements by a value of 2 ppm)

Output protection

Fully defibrillator protected and isolated

Multi-Function
Electrode (MFE)
Pads

Specifically designed adult anterior/posterior pre-gelled ZOLL MFE Pads and
Multi-Function stat-padz packaged in pairs

Waveform

Rectilinear Biphasic

Energy Selection

Adult Mode: Selectable at 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 15, 20, 30, 50, 70, 85, 100, 120, 150,
200 J
(Delivered into 50

 load).

Selected using controls on sternum paddles or unit front panel.

Charge Time

Less than 6 seconds with a new, fully charged battery (first 15 charges to 200 joules).
Depleted batteries result in a longer defibrillator charge time.

Less than 15 seconds when operating without a battery, using AC power alone at 90% of
the rated mains voltage.

Less than 15 seconds with a new, fully charged battery pack, depleted by up to 15 200
Joule discharges.

Less than 25 seconds from the initial power on, with a new, fully charged battery pack,
depleted by up to 15 200 Joule discharges.

Less than 25 seconds from the initial power on when operating without a battery, using AC
power alone at 90% of the rated mains voltage.

Less than 30 seconds from initiation of rhythm analysis (Semiautomatic mode) with a new,
fully charged battery pack, depleted by up to 15 200 Joule discharges.

Less than 30 seconds from initiation of rhythm analysis (Semiautomatic mode) when
operating without a battery, using AC power alone at 90% or greater of the rated mains
voltage.

Less than 40 seconds from initial power on (Semiautomatic mode) with a new, fully
charged battery pack, depleted by up to 15 200 Joule discharges.

Less than 40 seconds from initial power on (Semiautomatic mode) when operating without
a battery, using AC power alone at 90% or greater of the rated mains voltage.

Patient Impedance
Range

Minimum: 15 oms
Maximum: 250 ohms

Energy Display

Monitor display indicates both selected and delivered energy.

Synchronized Mode

Synchronizes defibrillator pulse to patient’s R-wave. SYNC message displayed on monitor.
Marker on display and recorder paper identifies R-wave discharge point. Meets the
DF-80:2003 requirement of 60ms maximum time delay between sync pulse and delivery of
energy, where the ECG is derived via an applied part, once the output has been activated.

Charge Controls

Control on apex paddle and on device front panel.