0 spo2 monitor test for spo2 option, Spo2 monitor test for spo2 option, 0 spo – ZOLL E Series Monitor Defibrillator Rev D User Manual
Page 39: Monitor test for spo, Option
E Series Service Manual
9650-1450-01 Rev. D
1-29
17.0 SpO
2
Monitor Test for SpO
2
Option
Tools Needed:
•
Masimo
Reusable Sensor.
•
Masimo
Patient Cable
•
Fluke Index Series 2 SpO
2
Simulator.
Test Setup:
1. Connect the universal cable to the MFC test plug.
2.
DO NOT connect the ECG cable to the simulator.
3.
Connect the Masimo
Patient Cable and attach the Masimo
sensor to the patient cable.
4.
Connect the Masimo
sensor to the finger simulation post.
5.
Place a fully charged battery into the battery well or connect to ac power (dc power, if equipped).
6.
Ensure that the SpO
2
Simulator is off.
Do this...
Verify that...
Pass/Fail
17.1
Turn the selector switch to MONITOR.
(For AED units, turn the selector switch to ON
and select Manual mode.)
The SpO
2
saturation percentage appears as a dashed line on the monitor.
o o
17.2
Wait ten seconds.
Turn on the SpO2 simulator.
Press the SIM
softkey on the Index SpO
2
Simulator, then press
the MAN softkey.
The unit displays the SpO
2
PULSE SEARCH message.
o o
17.3
Press the 02+ or 02- softkey of the simulator
until the SpO
2
output is at 98%.
The SpO
2
reading of 98 +/- 1% appears on the E Series monitor.
Note: You may need to wait up to 2 minutes for the information to appear on
the ZOLL display.
o o
17.4
Using the Index SpO
2
Simulator, press the
BPM+ or BPM- softkey until the heart rate is
230 BPM.
The SpO
2
simulator screen displays an SpO
2
rate of 230 BPM.
Note that you may need to wait up to 2 minutes for the information to appear on
the ZOLL display.
The E Series monitor displays an SpO
2
saturation of 96-100%.
The E Series monitor displays a heart
rate of 226-234 BPM.
o o