beautypg.com

Ransburg, Mma-570 direct/indirect charge - maintenance – Ransburg MMA-570 Direct_Ind Charge Atomizer A13366 User Manual

Page 66

background image

Interior/Exterior Shaping Air Manifold Removal

Reassembly
(Lightly lubricate all O-rings prior to assembling.)

Carefully install the inner shaping air ring onto the

turbine threads. Tighten in a clockwise direction

until it seats against the turbine. Tighten set

screw to 5 lbs•in (0.564 Nm) torque to prevent

shaping air ring/manifold from rotating. Do not

over-tighten! If replacing the solvent tube, install

into the atomizer body first and tighten with a 3/16”

end-wrench. Before installing the other end into

the inner shaping air ring, check the position of

the 1/4-20 threaded hole. If it is less than 180°

from the fitting installed in the atomizer body, you

must install a loop (as shown in “Installing Loop

In Tubing” figure) to prevent tube from becoming

pinched when outer shaping air ring is installed.

Do not kink the tube when installing loop (see

“Interior/Exterior Shaping Air Manifold Removal”

figure).

Turbine Removal/Replacement
Removal

Remove the turbine retaining ring by using the

spanner wrench (A12088-00), turning the turbine

retaining ring in a counter-clockwise direction. Pull

the turbine out while rocking it from side to side.

Turbine Removal

Replacement

Apply a light coating of O-ring lubricant to all the

O-rings and the threads of the turbine and turbine

retaining ring prior to assembly. Push the turbine

down into the cavity in the atomizer body. Align the

mark on the turbine with the mark on the atomizer

body. Install the turbine retaining ring and o-ring

by hand. Use the spanner wrench to tighten an

additional 1/8-1/4 turn. (Lightly lubricate O-ring

with petroleum jelly.) Check centering of fluid tube.

If fluid tube is centered, the turbine is fully seated.

If not, check tightness with spanner wrench. If

tube is not centered, again remove turbine and

check for causes, such as an O-ring fell off, fiber

optic not fully installed, foreign material on seating

surface, etc. Reinstall and recheck tube centering.

Fluid Tip Removal / Replacement

Removal

To remove the fluid tips, use the tip/tube removal

tool (A11229-00). Insert the tool over the tip and

engage the four (4) prongs of the tool into the four

(4) slots in the tips (see “Fluid Tip Removal” figure).

Fluid Tip Removal

†

To remove, turn the tip CLOCKWISE.

The thread on the tip is left hand.

NOTE

ENGAGE PRONGS

INTO SLOTS

A11229-00 TOOL

FLUID TIP

MMA-570 Direct/Indirect Charge - Maintenance

Ransburg

LN-9279-13

63