Digital In

The Digital I/O Ports are located on the

lower first block of the front panel. The

first port of the block is the Digital ground

(Dgnd). As Inputs, the digital signal has

an internal pull up resistor of 47K Ohms.

This allows a switch closure to ground to

activate the input. It can also be driven

using normal logic levels.

INPUTS

36

Port(s)

Defines which port is currently being

measured and will be used as the Task

Value.

Mode

Defines the current mode of the Digital Input. Digital Inputs have four modes of operation: Level,

Frequency, Event Counter, and Quad Counter. Level and Event Counter Modes allow the Edge

Triggered (Event-driven)Type to be used.

Level

The default mode for a Digital Input is

Level. This Digital Value gives the current

state of the digital I/O pin. As an Input, the

value will be a 1 or a 0 meaning “On” or

“Off” respectively. If the pin value is a 1,

the input voltage will be approximately 3.5

volts or higher, (normally 5.0 volts). If the

pin value is a 0, the input voltage will be

approximately 0.8 volts or less, (normally

0.0 volts). The range between 0.8 and 3.5

is not defined and may be either a 1 or a 0.

This Mode could be used to test the status

of an external device, such as a motor. The

digital I/O pin would be used as the

source for data; a 1 would be measured if the motor was on, a 0 would be measured if the motor

was off.