MAINTENANCE & TROUBLE SHOOTING

128

Sustained operation of the System 5000

is generally maintenance-free. There are no

desiccants or serviceable parts to maintain.

The touchscreen interface, however, can

be calibrated for finer precision when

necessary.

Maintenance

To re-calibrate the touchscreen, follow the onscreen prompts by pressing the center of the

crosshair displayed on the screen. A finer precision can be achieved by pressing each crosshair

for a few seconds, rather than a quick press. This allows more samples to be taken and a more

accurate averaging to take place.

Touchscreen Calibration

The touchscreen calibration menu can

be accessed by holding in the View

Status button while applying power to

the System 5000

. Continue to hold the

View Status button until the first

System 5000

splash screen appears.

Within a few seconds, a simple

touchscreen calibration menu will

appear.

When completed, the newly calibrated

settings can be tested.

If finer calibration settings are still

required, the calibration can be retried

by pressing the Retry Calibration button.

If the new settings are acceptable, press

the Exit and Restart button to reboot

the System 5000

with the new

calibration settings.