beautypg.com

How to test and adjust main control valve relief – Gradall 534B (9103-1390) Service Manual User Manual

Page 82

background image

TESTING HYDRAULIC RELIEF

VALVE PRESSURES

The reason for testing the main relief valves is to
assure that the relief valves open and to limit the
pressure in the lines at that time.

A. The engine should be turning at full governed

rpm.

B. The hydraulic oil should be warmed up to

operating temperature.

C. The test gage or gages should be installed in the

circuit being tested.

D. The circuit function should be stalled by either

running the cylinder to the end of its stroke or by
stalling the motor action.

E. With the control levers actuated, a gage reading

is taken. The gage indicates the setting of the
respective relief valve.

Circuit relief valves are used in some circuits to
protect Gradall Material Handler from external
forces.

Other relief valves are installed in the circuit to limit
the pressures for that circuit or to control flow
during specific actions.

NOTE

Park Material Handler on level ground with

forks on ground. Shut off engine, set parking
brake and remove ignition key. Chock the
wheels belore pertorming any servicing.

ADJUSTING RELIEF VALVES

All of the relief valves can be adjusted using an allen
wrench or a screwdriver. Loosen the lock nut, and
turn the adjusting screw clockwise to increase the
prssures. Turning it counter-clockwise will
decrease the pressures. The valve adjustment is
sensitive, so adjust in smalll moves, like 1/8 to 1/4
turn at a time. Tighten lock nut after adjustment.

HOW TO TEST AND ADJUST

MAIN CONTROL VALVE RELIEF

Set pressures with oil at operating temperature and
engine at 2800 rpm.

1. Install gage in test port on tandem pump out-

board tube. (C)

2. Retract boom fully and hold control lever full

to left.

3. Gage should read 3000 psi.

4. Adjust, reseal or replace relief valve, as

necessary.

5. Remove gage and recap test port.

If adjustment of the relief valve will not provide
desired pressure readings:

1. Replace pressure gage.

2. Test or examine other components in the circiut
being tested.

3. Replace defective component.

5