beautypg.com

Configuration diagrams, Full-bridge – Measurement Computing StrainBook/616 User Manual

Page 72

background image

Configuration Diagrams

Full-Bridge Configurations

The full-bridge has four strain-variable elements and requires no bridge completion components. Quarter and

half-bridge resistors may be left installed. Any bridge resistance from 60 to 1000 ohms can be accommodated.

Full-Bridge (+), Any Resistance from 60 to 1000 Ohms

IA

1

8

9

7

5

4

3

6

2

External
Bridge

DB-9
Input

Internal Bridge Completion

RA

RB

RC

RD

RE

RF

RG

RH

1

2

3

-Sense

+Sense
+Excitation

-Excitation

JN01

Switches
accessed
through
software

In this connection, excitation voltage is

regulated at the connector. This

configuration should only be used for

short cable lengths. Output polarity may

be altered by interchanging the (+) and (-)

input or by selecting the software invert

function.

Full-Bridge (+), with Remote Sense

IA

External
Bridge

DB-9
Input

Internal Bridge Completion

RA

RB

RC

RD

RE

RF

RG

RH

1

2

3

-Sense

+Sense
+Excitation

-Excitation

JN01

Switches
accessed
through
software

1

8

9

7

5

4

3

6

2

In this connection, excitation voltage is

regulated at the strain gage.

This eliminates errors due to cable

losses and is the preferred connection

for longer cables.


Full-Bridge (+), with B, D, or F Shunt

IA

External
Bridge

DB-9
Input

Internal Bridge Completion

RA

RB

RC

RD

RE

RF

RG

RH

1

2

3

-Sense

+Sense
+Excitation

-Excitation

JN01

Switches
accessed
through
software

1

8

9

7

5

4

3

6

2

The B, D, or F shunt resistor may be

software selected when installed as

shown. Output polarity during shunt

calibration will be automatically

corrected by software. The shunt resistor

value will typically be different for each

value of bridge resistance.

6-8 Bridge Configurations

929494

StrainBook/616 User’s Manual

This manual is related to the following products: