The Five Methods of Sensor Calibration

Before proceeding with calibration, remember to enter your password. The password

must be entered before channel parameters can be changed.
Unless all of the parameters (for each channel to be calibrated) are accurately entered

into the spreadsheet, the calibration will produce incorrect results.


The sensor calibration program uses five methods of calibration:

o

2-Point Manual

o

2-Point Automatic

o

Shunt (Internal)

o

2-Point Shunt (External)

o

Name Plate (for load cell or transducer)

Table of Calibration Methods and Required Parameters

Excit-

ation

Gage

Factor

Sensi-

tivity

Shunt
Ohms

Bridge

Ohms

Full

Load

Max.

Load

Quie-
scent

Pt-1

(mV)

Pt-1

(Units)

Pt-2

(mV)

Pt-2

(Units)

2 Point
Manual

9

9

9

9

9

2 Point
Auto

9

9

9

9

9

9

9

Shunt
(Internal)

9

9

9

9

9

9

9

2 Point
Shunt
(External)

9

9

9

9

9

9

Nameplate

9

9

9

9

9


To use any of these calibration methods, enter the appropriate values into the required spreadsheet columns

of the Sensor Calibration window, as listed above, and click on the Calibrate Enabled Channels button on

the toolbar.

In 2-Point Manual calibration, a message box prompts you to apply the first load. When

prompted, apply the load and click the OK button. A second message box will prompt

you to apply the second load. When prompted, apply the second load and click OK.

Saving a Calibration File.

After calibrating the enabled channels, a message box asks if you want to save

the changes. Click on the Yes button to save the calibration and a dialog box will appear. If you choose not

to save the changes at this time, another message will appear asking if you want to save the changes when

you click on the Return to WaveView button on the tool bar. Click on the Yes button to save these changes

and a dialog box will appear. The most recently saved calibration file will be recorded in the

WAVEVIEW.CFG

default configuration file and will be loaded into WaveView whenever a new session is

started. The current configuration can also be saved from the toolbar or File menu item.

4-14 Software