Synchronizing Multiple StrainBooks

StrainBooks can be synchronized to other StrainBooks, WBK40, or

WBK41 modules via SYNC ports. Each unit has two identical SYNC

ports located on the rear panel. Either [or both] SYNC ports can be

used to connect to the other devices via SYNC cables CA-74-1 or

CA-74-5 (1 ft. or 5 ft. cables, respectively). A StrainBook can be

scan-synchronized and triggered from any other StrainBook that is

connected to it via a SYNC cable.
The diagram shows how a StrainBook can input or output synchronization timing and trigger events on

either SYNC connector. A more detailed block diagram follows shortly.

Reference Note: The SYNC features are programmable via software. For detailed

information see the Using Multiple Devices and the daqAdcSetClockSource sections of

the Programmer’s Manual (p/n 1008-0901). A PDF version of the document is included on

the data acquisition CD.

PDF NOTE:

During software installation, Adobe

®

PDF versions of user manuals automatically install onto

your hard drive as a part of product support. The default location is in the Programs group,

which can be accessed from the Windows Desktop. Refer to the PDF documentation for

details regarding both hardware and software.

Synchronization Model

The synchronization model illustrates the SYNC function available with StrainBook/616.

Reference Note:

Chapter 10, System Expansion, includes a discussion of how up to 4 StrainBooks can be

synchronized.