External Shunt Calibration (2 Pt Shunt)

This calibration method requires the use of a WBK16/LC option, which is discussed elsewhere in the document.

The method supports the use of an external shunt resistor. The resistor is shunted across the gage using the (RF)

switch in the StrainBook or WBK16. The method appears simply as "Shunt" in the list of calibration methods

when either “Load Cell” or “Transducer” is selected as the sensor type.

Note: The WBK16/LC provides a non-committed dry contact on two pins of a single StrainBook or WBK16

channel connector. The WBK16/LC can be used for virtually all single value shunt calibration requirements,

some of which are not possible with the internal FET/analog switch provisions in the standard channel

configurations.

WBK16/LC

CAUTION

Make the DB9 pin # 2 connection to the “low-side” or the “high-side” of the

circuit. BUT NEVER TO BOTH! Doing so will create a short circuit that could

damage equipment.

External Shunt Resistor, Low-Side Connection

(see chapter 6 in regard to making a “High-side” connection)

Full Bridge (+) with External Shunt Resistor

DB9 Pin 2 Connected to the Low-Side

Reference Note:

The WBK16/LC Load Cell Shunt Cal internal option board is detailed in chapter 6.

That chapter includes an alternate schematic for connecting Pin 2 to the High-side.