StrainBook and WBK16 Daisy-Chain

StrainBook System with two WBK16 Modules

* Note that earlier TR-40U units have a limit of 2.7 A @ 15 VDC. Those supplies still provide adequate power for StrainBook and WBK16.

TR-40U

3.3 Amp limit

@ 15 VDC *

WBK16



WBK16




StrainBook/616

CA-129

CA-150

CA-242 or

CA-242-7

Device Interface Side Channel Input Side

The illustrated system consists on one StrainBook/616 and two WBK16 Strain Gage Modules. An

explanation of connections follows.
Power – A separate TR-40U power supply is being used for each of the three devices. The DIN5 cable

end of the TR-40U connects POWER IN DIN5 connector on the device. The other end of the TR-40U

connects to an AC outlet (90 to 264 VAC).
Ethernet – The CA-242 [or CA-242-7] Ethernet patch cable connects the StrainBook to the host PC’s

Ethernet port or to a network Ethernet hub.
Expansion Control – The CA-129 is a 15 pin connector cable used for Expansion Control.

o

One CA-129 connects StrainBook’s EXPANSION CONTROL port to the first WBK16’s

EXPANSION CONTROL IN port.

o

A second CA-129 runs from the first WBK16’s EXPANSION CONTROL OUT to the second

WBK16’s EXPANSION CONTROL IN.

Expansion Signals – The CA-150 is a BNC-to-BNC cable used to carry Expansion Signals. The

signals go to the StrainBook’s Analog-to-Digital Converter (ADC).

o

One CA-150 goes from the upper WBK’s EXPANSION SIGNAL OUT (BNC connector) to

the lower WBK16’s EXPANSION SIGNAL IN.

o

A second CA-150 connects the lower WBK16’s EXPANSION SIGNAL OUT to the

StrainBook’s EXPANSION SIGNAL IN (BNC connector).

Unit

Channel

#

StrainBook/616 1-8

1

st

WBK16

9-16

2

nd

WBK16

17-24

3

rd

WBK16

25-32

4

th

WBK16

33-40

5

th

WBK16

41-48

6

th

WBK16

49-56

How Analog Input Channel Numbers are Determined
The analog input channel numbers are determined by the order in which the WBK

modules are connected to the StrainBook.

Channels 1 through 8 are the StrainBook/616 channel inputs (via DB9).

Channels 9 through 16 are located on the first expansion WBK16 unit,

i.e., the module connected directly to the StrainBook.

For each additional WBK16 channel numbers are added in a group of 8

as indicated in the table at the right.


Note
: No more than seven WBK16 Expansion Modules can be attached to one

StrainBook/616.

7

th

WBK16

57-64

10-4 System Expansion

968594

StrainBook/616 User’s Manual