beautypg.com

AMETEK i-iX Series II User Manual

Page 96

background image

User Manual

California Instruments

96

i Series II / iX Series II / iM Series II

WAVE GROUP

[iX Series only] Sets the user defined waveform group that will
be loaded at power on. Available groups are 0, 1, 2 and 3. Each
group can contain up to 50 user-defined waveforms. A
waveform group can only be loaded at power up. To change
groups, you must change this field to the desired new group
and cycle the power to the AC source or issue a *RST
command over one bus.

CLOCK MODE

Sets the clock source used at power up. Available settings are
Stand Alone (STAND), MASTER, and SLAVE.

NO. OUTPUT

[iX Series only] Sets the phase mode at power up. Available
options are determined by the presence of the phase mode
option. If the phase mode option is installed, available options
are single phase mode (ONE), three phase mode (THREE) or
last used phase mode (LAST). If the phase mode option is not
installed, this field is either fiXed to ONE or THREE.

VOLT ALC

Determines ALC mode at power on. The ALC mode adjusts the
output voltage based on internal voltage measurement system
and provides enhanced output regulation and accuracy.
Available settings are ON, OFF or REG.