Note: The Fluke 8506A Digital Multimeter must be used for the following calibration. The

8506A must be set to the AC HI ACCUR mode for all AC measurements.

AC Volt Full-scale:

1. Select the AC mode of operation.

1. Program the output to the 300 VAC range. Close the output relay.

2. Program the output to 240 volts and 60 Hz.

3. Go to the MEASUREMENT CALIBRATION screen.

4. Enter the actual AC output voltage for the VOLT FS parameter and press the ENTER

key.

5. Allow the controller to complete the calibration cycle. This may take several minutes.

6. When completed, the new cal coefficient will be displayed.

7. Press the SET key to save the new calibration coefficients.

AC Current Full-scale:

2. Select the AC mode of operation.

3. Apply a load to the output. Refer to Table 6-1.

4. Program the output to 135 VAC and 60 Hz on the 150 Volt AC range.

5. Observe the actual output current and enter this value for the CURR FS parameter.

Press the ENTER key.

6. Allow the controller to complete the calibration cycle. This may take several minutes.

7. When completed, the new cal coefficient will be displayed.

8. Press the SET key to save the new calibration coefficients.

DC Volt Full-scale:

1. Select the DC mode of operation.

2. Program the output to the 400 VDC range. Close the output relay.

3. Program the output to 320 volts DC.

4. Go to the MEASUREMENT CALIBRATION screen.

5. Enter the actual DC output voltage for the VOLT FS parameter and press the ENTER

key.

6. Allow the controller to complete the calibration cycle. This may take several minutes.

7. When completed, the new cal coefficient will be displayed.

8. Press the SET key to save the new calibration coefficients.

DC Current Full-scale:

1. Select the DC mode of operation.

2. Apply a load to the output. Refer to Table 6-1.

3. Program the output to 120 VDC on the 200 Volt DC range.

4. Observe the actual output current and enter this value for the CURR FS parameter.

Press the ENTER key.

5. Allow the controller to complete the calibration cycle. This may take several minutes.