6. When completed, the new cal coefficient will be displayed.

7. Press the SET key to save the new calibration coefficients.

TITLE

PROGRAM / LOAD PARAMETERS

PARAMETER

ADJUST TO

AC Volt Full-scale

240 VAC, 60 Hz

VOLT FS

V

ac

AC Current Full-scale

150 Range, 135 VAC, full load

CURR FS

I

ac

DC Volt Full-scale

320 VDC

VOLT FS

V

dc

DC Current Full-scale

200 Range, 120 VDC, full load

CURR FS

I

dc

Table 6-2: Measurement Calibration Table

For a multi-phase power system that uses one controller, 9003i/iX, 10002i/iX, 15003i/iX or
30003i/iX, repeat paragraph 6.3 for each phase. Move the external test equipment to the phase
that is being calibrated. Refer to Figure 6-1.

While viewing the calibration screen, press the PHASE key to select the respective phase.