beautypg.com

Specifications – AMETEK i-iX Series II User Manual

Page 15

background image

User Manual

California Instruments

i Series II / iX Series II / iM Series II

15

2. Specifications

All specifications are for a single i or iX Series II unit and 25

±

5

°

C sine wave output with a

resistive load unless noted otherwise.

2.1 Electrical

2.1.1 Input

Parameter

3001i / iX / iM

5001i / iX / iM

Line Voltage:

208-240

±

10% VAC, single phase 208-240 V

LL

±

10%, (Standard)

400-440 V

LL

±

10%, (-400)

400-480 V

LL

±

10%, (-400)

3 phase, 3 wire + ground

Line VA:

5000VA

8000VA

Line Current:

25 A RMS max. (Per Box)

23 A RMS max. at 208-240 VAC

12 A RMS max. at 400-440 VAC
and 400-480 VAC (Per Box)

Line Frequency:

50-60 Hz

±

10%

Efficiency:

80% (typical) depending on line and load

Power Factor:

0.7 (typical)

0.9 (typical)

Inrush Current:

100 A

pk

for 100

µ

s

100 A

pk

for 100

µ

s at 208-240V

50 A

pk

for 100

µ

s at 400-440 VAC

and 400-480 VAC

Hold-Up Time:

15 ms

Isolation Voltage:

2200 VAC input to output

1350 VAC input to chassis