beautypg.com

AMETEK i-iX Series II User Manual

Page 197

background image

User Manual

California Instruments

i Series II / iX Series II / iM Series II

197

(limited by 5001iX / 15003iX)

at 0% U

NOM

N/A

N/A

Bypass Mode Engaged:

Maximum Current

37 A

RMS

37 A

RMS

/phase

9.4.4.3 Controls/Indicators

Control / Indicator

Comments

ON/OFF toggle switch

70 / 80 Tap Select toggle switch

(EOS1, second generation only).

Power On lamp

Active lamp

Fault lamp

9.4.4.4 Mechanical

Dimensions

7"H x 19"W x 24"D

178 mm H x 483 mm W x 610 mm D

Weight

EOS-1

70 lbs, 32 kg

EOS-3

160 lbs, 73 kg

Connectors:

Input mains

IEC 320

Source

Compression

Load

Compression

Option In/Out

9 pin D-subminiature

Chassis Slide

General Devices C300S-120-B308

9.4.4.5 Environmental

Temperature

0 to 40 °C

Relative Humidity

80% max. to 31 °C

derate linearly to 50% max. at 40°C

Altitude

2000 m max.

Installation Over voltage Category

II

Pollution Degree

2

Equipment Class

I

Usage

Indoors