AMETEK i-iX Series II User Manual
Page 16

User Manual
California Instruments
16
i Series II / iX Series II / iM Series II
2.1.2 Output
(ALL SPECIFICATIONS ARE FOR AC AND DC UNLESS NOTED OTHERWISE)
Output Parameter
i / iM Series
iX Series
Modes:
AC, DC
AC, DC, AC+DC
Voltage:
Ranges (L-N):
AC Mode
Low:
0 - 150 VAC
High
1
: 0 - 300 VAC
DC Mode
Low:
0 - 200 VDC
High:
0 - 400 VDC
AC+DC Mode
iX Models only.
AC:
Low: 0 - 150 V / High: 0 - 300 V
(See footnote 1)
DC
Offset: Low 0 - 150 V / High; 0 - 225 V
Programming
Resolution:
AC Mode
0.1 V
DC Mode
0.1 V
AC+DC Mode
AC:
0.1 V
DC Offset:
0.1 V
Voltage Accuracy:
AC mode
±
0.5% of range, 16 to 400
Hz.
±
0.5% of range, 16 to 400 Hz.
DC mode
±
0.5% of range
±
0.5% of range
Voltage Distortion
2
:
(linear load)
1% max THD at 50/60 Hz
2% max THD at 400 Hz
3% max THD at 1000 Hz
1% max THD at 50/60 Hz
2% max THD at 400 Hz
3% max THD at 1000 Hz
Load Regulation (% FS
Vrange):
ALC on
±
0.2%
±
0.2%
ALC off
±
0.5% DC to 100 Hz
±
2.2% to 1000 Hz (Low
range)
±
0.6% to 1000 Hz (High
range)
±
0.5% DC to 100 Hz.
±
2.2% to 1000 Hz (Low range)
±
0.6% to 1000 Hz (High range)
Line Regulation:
0.1% for 10% input line
change
0.1% for 10% input line change
Power: (per phase, either range, at full scale voltage)
1
Maximum RMS voltage in high voltage range is a function of programmed frequency. For frequencies above 500
Hz, the maximum available Vrms = 1.5 E+4 / F. See Figure 2-9 for V-F rating.
2
The distortion specification for the 3001i and iX is valid for an input voltage range of 197-264 V.