RUN ALL

The RUN ALL selection will cause the following automated test sequence suggested by the
standard to be run:

Step

Output in

% of U

T

No of Cycles

Start angle

(degrees)

Repeat #

times

Delay

between

repeats (s)

1

0

0.5

0

3

10

2

0

0.5

180

3

10

3

0

1

0,45,90

3 at diff ø

10

4

0

5

45,90,135

3 at diff ø

10

5

0

10

90,135,180

3 at diff ø

10

6

0

25

180,225,270

3 at diff ø

10

7

0

50

270,315,0

3 at diff ø

10

8

40

0.5

0

3

10

9

40

0.5

180

3

10

10

40

1

0,45,90

3 at diff ø

10

11

40

5

45,90,135

3 at diff ø

10

12

40

10

90,135,180

3 at diff ø

10

13

40

25

180,225,270

3 at diff ø

10

14

40

50

270,315,0

3 at diff ø

10

15

70

0.5

0

3

10

16

70

0.5

180

3

10

17

70

1

0,45,90

3 at diff ø

10

18

70

5

45,90,135

3 at diff ø

10

19

70

10

90,135,180

3 at diff ø

10

20

70

25

180,225,270

3 at diff ø

10

21

70

50

270,315,0

3 at diff ø

10

Table 9-19: Dips and Interruptions Tests Performed During RUN ALL

The user can change the NOMINAL Ut voltage for this. The RUN ALL Command line will change
to ABORT during the test. Selecting ABORT and pressing the ENTER key will terminate the test
at any time and the output voltage will return to the nominal value.

RUN SINGLE

RUN SINGLE command will run a single test once. The Dip or Interrupt test is defined by the
DIP TO, NO CYCLES, and START ANGLE parameters. These parameters must be set before
starting the test. The following is a description of these parameters.

DIP TO:

The dip voltage level as a percentage of the nominal voltage.

NO CYCLES:

The dip duration in cycles.

START ANGLE:

The start phase angle of the dip

Note: After each individual run, a 10 second delay is inserted.