beautypg.com

AMETEK i-iX Series II User Manual

Page 191

background image

User Manual

California Instruments

i Series II / iX Series II / iM Series II

191

Figure 9-28: IEC 61000-4-13 Test Flowchart Class 1 and 2


Start - Class 1/2

Determine appropriate test

configuration

Perform 8.2.1 Test

"Harmonic combination"

Flatcurve and Overswing

Any functional

anomalies?

No

No

Class 2

required?

Meister

curve required?

Any functional

anomalies?

Any functional

anomalies?

Any functional

anomalies?

Test Completed

FAIL

Record

Results

Test Completed

PASS

Record

Results

Yes

Yes

No

Perform 8.2.2 Test

"Frequency Sweep"

Perform 8.2.4 Test

"Meister Curve"

Perform 8.2.3 Test

"Individual Harmonics/

Interharmonics"

Yes

Yes

No

Yes

Yes

Yes

No