beautypg.com

AMETEK i-iX Series II User Manual

Page 238

background image

User Manual

California Instruments

238

i Series II / iX Series II / iM Series II

9.8.1.6 Revision Selection

The default Revision is E. Revisions supported are D and E. The Revision can be changed from
the front panel. Scroll to the REVISION entry using the up and down cursor keys (see Figure
9-49). Use the shuttle to change the selection.

9.8.1.7 AC MODE TESTS

9.8.1.7.1 Steady StateTest

Scroll to the STEADY STATE AC entry using the up and down cursor keys. Press the ENTER
key to select the STEADY STATE screens. The screen will appear as shown in Figure 9-50.

Figure 9-50: Steady State Menu

The MIL704 Steady state 1 and Steady state 2 screens have the following tests:

1

VOLTAGE

2

FREQUENCY

3

VOLT UNBALANCE

4

PHASE DIFFERENCE

5

VOLT MODULATION

6

FREQ MODULATION

7

DISTORTION

The above tests can be selected by scrolling to the highlighted selection using the up and down
cursor keys and the ENTER key to start the selected test.

VOLTAGE

This test will change the output voltage in the following sequence:

108V for 1 minute.

118V for 1 minute.

115V for 1 minute.

The

key (backspace) will terminate the test at any time.

FREQUENCY

This test will change the output frequency in the following sequence:

393Hz for 1 minute.

407Hz for 1 minute.

400Hz for 1 minute.

The

key (backspace) will terminate the test at any time.

VOLT UNBALANCE

This test will change the output voltage for the selected phase only in the following sequence: