Figure 9-29: IEC 61000-4-13 Test Flowchart Class 3

Start - Class 3

Determine appropriate test

configuration

Perform 8.2.1 Test

"Harmonic combination"

Flatcurve and Overswing

Any functional

anomalies?

Any functional

anomalies?

Any functional

anomalies?

Any functional

anomalies?

Test Completed

FAIL

Record

Results

Test Completed

PASS

Record Results

Yes

Perform 8.2.2 Test

"Frequency Sweep"

Perform 8.2.4 Test

"Meister Curve"

Perform 8.2.3 Test

"Individual Harmonics/

Interharmonics"

Yes

Yes

Yes

No

No

Any functional

anomalies?

Yes

Perform 8.2.3 Test

Only table 1. Odd

non-multiple of

3 harmonics