9.6.6.5 Control Configuration

Manual versus programmable control of the OMNI is configured using a two pole DIP switch on
the DC Supply/Relay Board assembly inside the OMNI chassis. To reconfigure the control
feature, perform the following steps:

WARNING: Disconnect all sources of power to the OMNI before performing any of these steps.
Hazardous voltages may be present internally from the AC mains and from AC power source
connected to the OMNI rear panel.

1.

Disconnect all sources of power to the OMNI.

2.

Remove the OMNI top cover.

3.

Configure S1 DIP switch setting according to table:

MODE

S1 POSITION 1

S1 POSITION 2

Manual

OFF

ON

Programmable

ON

OFF


Note: Position 1 is toward the OMNI front panel, position 2 is toward the OMNI rear panel.

4.

Replace the top cover.

5.

Reconnect power to the OMNI