6.3 Routine Measurement Calibration

Note: The Fluke 8506A Digital Multimeter must be used for the following calibration. The

8506A must be set to the AC HI ACCUR mode for all AC measurements.

Connect the test equipment to the power source as shown in Figure 6-2. If the power system is
a multi-phase system with one controller, the DVM for calibrating the measurement voltage
should always be connected to the Remote Sense connector (TB3) on the Phase A power
source.

The shunt must be connected to the power source as shown in Figure 6-1. If the Current
measurement can’t be successfully performed, adjust the Current Measurement Pot. This
adjustment is described in the Non-routine Calibration section of this manual. If the DC current
measurement displays more than 70 counts on the display, perform the non-routine current
monitor adjustment.

Connect the load to the output. Recommend current shunt values are 0.010 ohm for
3001/5001/9003/15003 and 0.001 ohm for 10001, 15001.

To calibrate all measurement functions, the desired value for the measurement value of current
or voltage must be entered for the corresponding calibration value. Make the indicated
adjustments by typing in the desired display value. This should be the value indicated by the
external DVM. If a 10 milliohm current shunt is used for current, 300 mV represents 30 amps.

The Measurement Calibration Table is a summary of the measurement calibration procedure.
The following text is a detailed explanation of the procedure.

PARAMETER

POWER SYSTEM

3001 / 9003

5001 / 15003

10001 / 30003

15001

AC Current Full-Scale

6.0

, 3 KW

3.6

, 5 KW

1.8

, 10 KW

1.3

, 15 KW

DC Current Full-Scale

13.5

, 1.5 KW

4.8

, 3.5 KW

2.4

, 7 KW

1.6

, 10.5 KW

Table 6-1: Calibration Load For Each Phase