4.2.8 ADVANCE MEAS. Menu [iX Series only]

This entry in the MENU 2 screen displays the HARMONICS/TRACE ANALYSIS screen which is
covered in section 0. This field can be used in lieu of the MEAS key to directly bring up the
advanced measurements screens.

4.2.9 APPLICATIONS Menu

Figure 4-18: APPLICATIONS menu

The APPLICATIONS menu provides access to the optional application specific pre-programmed
test sequences. Since these test sequences are optional, this menu may have no choices if
none of the options are installed. The following entries may be found in the APPLICATIONS
menu:

Entry

Description

MIL-STD 704

Test sequence for MIL standard 704 AC and DC tests.

OMNI OPTION

This entry selects the OMNI control menu. The OMNI option
provides a lumped reference impedance for use in IEC 1000-3-
3 Flicker test applications. The iX Series also provides
programmable impedance, so, in most cases, the OMNI option
is not required. For the i Series, the OMNI impedance is
available as an option. If the OMNI option is installed, this
screen allows the impedance to be ENGAGED or BYPASSED.

RTCA/DO-160

Test sequence for RTCA DO160 commercial aviation AC and
DC tests.

IEC-1000-4-11

Test sequences for IEC 1000-4-11 Voltage Dips and Variations
test standard.

IEC-1000-4-13

Test sequence for IEC 1000-4-13 Harmonics and Inter
harmonics test standard.

WH METER

Watt Hour meter measurement option. Tracks energy usage
over a period of time and calculates Watt Hours used.