AMETEK i-iX Series II User Manual
Page 181

User Manual
i Series II / iX Series II / iM Series II
RUN SINGLE
RUN SINGLE command will run the test once. The Variation test is defined by the REDUCE TO,
FALL TIME, HOLD TIME and RISE TIME parameters. These parameters must be set before
starting the test. The following is a description of these
REDUCE
TO:
The lowest voltage level as a percentage of the nominal voltage. Thus, 0% is 0
Volts. 100% is full nominal voltage.
FALL TIME:
The time in seconds it will take the output to reach the REDUCE TO voltage.
Values must be entered in
voltage drops instead of voltage sweeps to support Edition 2.0 of the test
standard. This requires firmware 2.38 or higher however.
HOLD TIME:
The time in seconds the output will hold at the REDUCE TO volt
must be entered in seconds.
RISE TIME:
The time in seconds the output will reach the NOMINAL voltage from the
REDUCE TO voltage. Values must be entered in seconds.
The timing of voltage variations is different between editions 1.0 and 2.0 of
test standard. While the original standard used voltage ramps on both sides of the variation test
level, the Edition 2.0 standard calls out an abrupt drop to the test level, followed by a voltage
ramp back to the nominal voltage. Thus, f
seconds.
Furthermore, where under Edition 1.0 rise and fall times were specified in seconds, they are
specified in number of cycles in Edition 2.0. This is an effort to normalize both 50 Hz and 60 Hz
voltage variations.
Refer to Figure 9-17
RMS output under the different standard revisions.
Figure
Figure 9
i Series II / iX Series II / iM Series II
RUN SINGLE command will run the test once. The Variation test is defined by the REDUCE TO,
FALL TIME, HOLD TIME and RISE TIME parameters. These parameters must be set before
starting the test. The following is a description of these parameters.
The lowest voltage level as a percentage of the nominal voltage. Thus, 0% is 0
Volts. 100% is full nominal voltage.
The time in seconds it will take the output to reach the REDUCE TO voltage.
Values must be entered in seconds. A 0.000 sec time may be used for abrupt
voltage drops instead of voltage sweeps to support Edition 2.0 of the test
standard. This requires firmware 2.38 or higher however.
The time in seconds the output will hold at the REDUCE TO volt
must be entered in seconds.
The time in seconds the output will reach the NOMINAL voltage from the
REDUCE TO voltage. Values must be entered in seconds.
The timing of voltage variations is different between editions 1.0 and 2.0 of
test standard. While the original standard used voltage ramps on both sides of the variation test
level, the Edition 2.0 standard calls out an abrupt drop to the test level, followed by a voltage
ramp back to the nominal voltage. Thus, for Edition 2.0, the fall time must be set to 0.000
Furthermore, where under Edition 1.0 rise and fall times were specified in seconds, they are
specified in number of cycles in Edition 2.0. This is an effort to normalize both 50 Hz and 60 Hz
17 and Figure 9-18 for an illustration of how these parameters affect the V
RMS output under the different standard revisions.
Figure 9-17: EN 61000-4-11 Voltage Variation specification
9-18: EN 61000-4-11 Voltage Variation specification- Edition 2.0
California Instruments
181
RUN SINGLE command will run the test once. The Variation test is defined by the REDUCE TO,
FALL TIME, HOLD TIME and RISE TIME parameters. These parameters must be set before
parameters.
The lowest voltage level as a percentage of the nominal voltage. Thus, 0% is 0
The time in seconds it will take the output to reach the REDUCE TO voltage.
seconds. A 0.000 sec time may be used for abrupt
voltage drops instead of voltage sweeps to support Edition 2.0 of the test
standard. This requires firmware 2.38 or higher however.
The time in seconds the output will hold at the REDUCE TO voltage. Values
The time in seconds the output will reach the NOMINAL voltage from the
REDUCE TO voltage. Values must be entered in seconds.
The timing of voltage variations is different between editions 1.0 and 2.0 of the IEC 61000-4-11
test standard. While the original standard used voltage ramps on both sides of the variation test
level, the Edition 2.0 standard calls out an abrupt drop to the test level, followed by a voltage
or Edition 2.0, the fall time must be set to 0.000
Furthermore, where under Edition 1.0 rise and fall times were specified in seconds, they are
specified in number of cycles in Edition 2.0. This is an effort to normalize both 50 Hz and 60 Hz
for an illustration of how these parameters affect the V
11 Voltage Variation specification- Edition 1.0
Edition 2.0